
|
OXYGEN
ANALYZER
VISIT
OUR NEW SITE
Click
Here
Oxygen Analyzer
Sensor Types
Today's oxygen analyzers use
one of a several types of oxygen sensors. As industrial process
applications call for improved measurement accuracy and repeatability,
users of oxygen analyzers are also demanding oxygen analyzers that require
a minimum of maintenance and calibration. To this end, users of
oxygen analyzers are encouraged to evaluate the merits of a particular
oxygen sensor type in context to the application for which it is intended.
There is no one universal oxygen analyzer type.
The synoptic review of the various
gas phase oxygen sensors provided below should be used in conjunction
with information gathered from manufacturers of oxygen analyzers. This
combination will help to ensure the selection of the right sensor type
for the application under consideration.
Oxygen
Analyzer with Ambient Temperature Electrochemical Oxygen Sensors.
Electrochemical-Oxygen Analyzer:
The ambient temperature electrochemical
sensor, often referred to as a galvanic sensor, is typically a small,
partially sealed, cylindrical device (1-1/4” diameter by 0.75” height)
that contains two dissimilar electrodes immersed in an aqueous electrolyte,
commonly potassium hydroxide. As oxygen molecules diffuse through a semi-permeable
membrane installed on one side of the sensor, the oxygen molecules are
reduced at the cathode to form a positively charge hydroxyl ion. The hydroxyl
ion migrates to the sensor anode where an oxidation reaction takes place.
The resultant reduction/oxidation reaction generates an electrical current
proportional to the oxygen concentration in the sample gas. The current
generated is both measured and conditioned with external electronics and
displayed on a digital panel meter either in percent or parts per million
concentrations. With the advance in mechanical designs, refinements
in electrode materials, and enhanced electrolyte formulations, the galvanic
oxygen sensor provides extended life over earlier versions, and are recognized
for their accuracy in both the percent and traces oxygen ranges. Response
times have also been improved. A major limitation of ambient temperature
electrochemical sensors is their susceptibility to damage when used with
samples containing acid gas species such as hydrogen sulfide, hydrogen
chloride, sulfur dioxide, etc. Unless the offending gas constituent is
scrubbed prior to analysis, their presence will greatly shorten the life
of the sensor. The galvanic sensor is also susceptible to over pressurization.
For oxygen analyzer applications where the sample pressure is > 5 psig,
a pressure regulator or control valve is normally recommended.
Paramagnetic
Oxygen Analyzer
Within this category, the magnetodynamic or `dumbbell' type of design
is the predominate sensor type. Oxygen has a relatively high magnetic
susceptibility as compared to other gases such as nitrogen, helium, argon,
etc. and displays a paramagnetic behavior. The paramagnetic oxygen sensor
consists of a cylindrical shaped container inside of which is placed a
small glass dumbbell. The dumbbell is filled with an inert gas such
as nitrogen and suspended on a taut platinum wire within a non-uniform
magnetic field. The dumbbell is designed to move freely as it is suspended
from the wire. When a sample gas containing oxygen is processed through
the sensor, the oxygen molecules are attracted to the stronger of the
two magnetic fields. This causes a displacement of the dumbbell which
results in the dumbbell rotating. A precision optical system consisting
of a light source, photodiode, and amplifier circuit is used to measure
the degree of rotation of the dumbbell. In some paramagnetic oxygen sensor
designs, an opposing current is applied to restore the dumbbell to its
normal position. The current required to maintain the dumbbell in it normal
state is directly proportional to the partial pressure of oxygen and is
represented electronically in percent oxygen. There are design variations
associated with the various manufacturers of magnetodynamic paramagnetic
oxygen analyzer types. Also, other types of sensors have been developed
that use the susceptibility of oxygen to a magnetic field which include
the thermomagnetic or `magnetic wind' type and the magnetopneumatic sensor.
In general, paramagnetic oxygen sensors offer very good response
time characteristics and use no consumable parts, making sensor life,
under normal conditions, quite good. It also offers excellent precision
over a range of 1% to 100% oxygen. The magnetodynamic sensor is quite
delicate and is sensitive to vibration and/or position. Due to the loss
in measurement sensitivity, in general, the paramagnetic oxygen sensor
is not recommended for trace oxygen measurements. Other gases that exhibit
a magnetic susceptibility can produce sizeable measurement errors. Manufacturers
of the paramagnetic oxygen analyzer should provide details on these interfering
gases.
Polarographic
Oxygen Analyzer
The oxygen analyzer that features a polarographic oxygen sensor is often
referred to as a Clark Cell [J. L. Clark (1822- 1898)]. In this type of
sensor, both the anode (typically silver) and cathode (typically gold)
are immersed in an aqueous electrolyte of potassium chloride. The electrodes
are separated from the sample by a semi-permeable membrane that provides
the mechanism to diffuse oxygen into the sensor. The silver anode is typically
held at a potential of 0.8V (polarizing voltage) with respect to the gold
cathode. Molecular oxygen is consumed electrochemically with an accompanying
flow of electrical current directly proportional to the oxygen concentration
based on Faraday's law. The current output generated from the sensor is
measured and amplified electronically to provide a percent oxygen measurement.
One of the advantages of the polarographic oxygen sensor is that while
inoperative, there is no consumption of the electrode (anode). Storage
times are almost indefinite. Similar to the galvanic oxygen sensor, they
are not position sensitive. Because of the unique design of the polarographic
oxygen sensor, it is the sensor of choice for dissolved oxygen measurements
in liquids. For gas phase oxygen measurements, the polarographic oxygen
analyzer type is suitable for percent level oxygen measurements only.
The relatively high sensor replacement frequency is another potential
drawback, as is the issue of maintaining the sensor membrane and electrolyte.
A variant to the polarographic
Oxygen Analyzer is what some manufacturers refer to as as oxygen analyzer
that uses a non-depleting coulometric sensor where two similar electrodes
are immersed in an electrolyte consisting of potassium hydroxide. Typically,
an external EMF of 1.3 VDC is applied across both electrodes which acts
as the driving mechanism for reduction/oxidation reaction. The electrical
current resulting from this reaction is directly proportional to the oxygen
concentration in the sample gas. As is the case with other sensor types,
the signal derived from the sensor is amplified and conditioned prior
to displaying. Unlike the conventional polarographic oxygen sensor, this
type of sensor can be used for both percent and trace oxygen measurements.
However, unlike the zirconium oxide, one sensor cannot be used to measure
both high percentage levels as well as trace concentrations of oxygen.
One major advantage of this sensor type is its ability to measure parts
per billion levels of oxygen. The sensors are position sensitive and replacement
costs are quite expensive, in some cases, paralleling that of an entire
oxygen analyzer of another sensor type. They are not recommended for applications
where oxygen concentrations exceed 25%.
Zirconium
Oxide Oxygen Analyzer
The type of oxygen analyzer that uses this type of oxygen sensor is occasionally
referred to as the “high temperature” electrochemical sensor and is based
on the Nernst principle [W. H. Nernst (1864-1941)]. Zirconium oxide sensors
use a solid state electrolyte typically fabricated from zirconium oxide
stabilized with yttrium oxide. The zirconium oxide probe is plated on
opposing sides with platinum which serves as the sensor electrodes. For
a zirconium oxide sensor to operate properly, it must be heated to approximately
650 degrees Centigrade. At this temperature, on a molecular basis, the
zirconium lattice becomes porous, allowing the movement of oxygen ions
from a higher concentration of oxygen to a lower one, based on the partial
pressure of oxygen. To create this partial pressure differential, one
electrode is usually exposed to air (20.9% oxygen) while the other electrode
is exposed to the sample gas. The movement of oxygen ions across the zirconium
oxide produces a voltage between the two electrodes, the magnitude of
which is based on the oxygen partial pressure differential created by
the reference gas and sample gas. The zirconium oxide oxygen sensor exhibits
excellent response time characteristics. Another virtue is that the same
sensor can be used to measure 100% oxygen, as well as parts per billion
concentrations. Due to the high temperatures of operation, the life of
the sensor can be shortened by on/off operation. The coefficients of expansions
associated with the materials of construction are such that the constant
heating and cooling often causes “sensor fatigue”. A major limitation
of the zirconium oxide oxygen analyzer is their unsuitability for trace
oxygen measurements when reducing gases (hydrocarbons of any species,
hydrogen, and carbon monoxide) are present in the sample gas. At operating
temperatures of 650 degrees Centigrade, the reducing gases will react
with the oxygen, consuming it prior to measurement thus producing a lower
than actual oxygen reading. The magnitude of the error is proportional
to the concentration of reducing gas. The zirconium oxide oxygen
analyzer is the “defacto standard” for in-situ combustion control applications.
Other types of oxygen analyzer
types under development and in some cases being used for specific applications.
They include, but are not limited to, luminescence polarization, opto-chemical
sensors, laser gas sensors, et al. As these techniques are further developed
and improved, they may represent viable alternatives to the existing technologies
used in today's oxygen analyzer.
Key words-Oxygen Analyzer-Trace
Oxygen Analyzer -Percent
Oxygen Analyzer - PPM
Oxygen Analyzer - Percent
Oxygen Analyzer - Industrial
Oxygen Analyzer - Medical
Oxygen Analyzer - Diving
Oxygen Analyzer -Zirconium
|